The GF-D800 semi-automatic biochemical analyzer is popular among grassroots users for its simple operation, stable performance, accurate results, low failure rate, and economical and practical use. In the process of using the machine, the author strictly carried out maintenance procedures. However, after less than half a year of use, some failures sometimes occur, affecting the normal operation of the instrument. The author has used various semi-automatic biochemical analyzers for decades. After exploration, I summarized some troubleshooting methods for the reference of my peers.
1. Failure phenomenon The instrument determines that all other items are operating normally. It is a crash when measuring the urea item alone, which makes the urea item unable to be measured (enzymatic method).
2. Trouble-shooting â‘ The instrument enters the diagnostic program, and the wavelength, temperature, and light energy are measured, and everything is normal. â‘¡ Use the recovery program CD to recover the program again, the fault is not eliminated. â‘¢ Open the shell of the instrument, replace the light source lamp, and remove the mirror dust. â‘£ Restart the instrument and troubleshoot.
3. Discuss and understand the establishment of instrument maintenance SOP file, the instrument should be managed by a dedicated person. Regularly replace the light source lamp, pump tube, sample tube, remove instrument dust, and mechanical parts can be added with medical petrolatum to ensure the normal operation and service life of the machine. After each maintenance and repair, detailed records should be made to ensure that the instrument is in the best condition, greatly reducing the failure rate, extending the service life of the instrument, and improving work efficiency.
Touch Panel For Iphone,Touch Screen Digitizer Panel,Touch Screen Panel For Iphone,Phone Touch Screen
Shenzhen Xiangying touch photoelectric co., ltd. , https://www.starstp.com